dhi
Products/Software
 
BioPRYN® 
Pregnancy Blood
Test Kits
Agricultural Technology
a
a
a
a
 
Lab System Pages
Foss Simple CA83 Interface
Printing in Windows XP
Configuration Backup
Purge Procedures
Revised Check Sample Standards
Troubleshooting: Making a LSINFO diskette.
Lab Data Collection

Check Sample Standards

The Check Sample Standards configuration screen is accessed by pressing Alt+F7 in the data collection screen (Main Menu / Collect data from lab instruments).

Check Sample Configuration:

ID - A single character to identify this check sample.

ls

Name
- A name to identify this check sample.

Machine 1
and Machine 2 columns - The expected Fat, Protein, Lactose, SNF, SCC, and MUN values for each machine.

Tolerances - There are three tolerance columns: Dev (deviation), Rep (repeatability), and Prg (purging efficiency). The tolerance values are the same for both machines.

Dev - The Deviation column values assume 2 decimal places for Fat, Protein, Lactose, and SNF (4 = 0.04). For SCC and MUN the Dev value is a percentage (7 = SCC must be within 7% of the standard).

Rep - The Repeatability column values assume 2 decimal places for Fat, Protein, Lactose, and SNF (4 = 0.04). For SCC the Rep value is coefficient of variation (CV) with 3 assumed decimal places (40 = 0.040).

Prg - The Purging Efficiency column values are the percentage the measured Purging Efficiency may deviate from 100% (1 = allowed range is 99% to 101%).

Samples - Number of samples (drinks) to take before calculating results.

Repeat - Number of times to repeat this check (take the number of samples specified and calculate results) for each instance that this check is logged on. Typically this is set to one.

Check Sample Results Printout
C 1. 352 319 449 749 481 11.99
C 2. 349 321 452 749 486 12.12
C 3. 352 319 450 754 513 12.14
C 4. 351 319 447 753 486 12.24
05/25/04 12:14:00
Machine: 121
CheckID: A - Hourly Check
FAT AVG 351
1
(350 +/- 4)
IN
FAT REP 3
(<= 4)
IN
FAT PRG 100%
(+/- 1%)
IN
PRO AVG 320
0
(320 +/- 4)
IN
PRO REP 2
(<= 4)
IN
PRO PRG 100%
(+/- 1%)
IN
LAC AVG 450
1
(449)
 
LAC REP 5
   
LAC PRG 100%
   
SNF AVG 751
1
(750)
 
SNF REP 5
   
SNF PRG 100%
   
SCC AVG 492
-15%
(578 +/- 7%)
Out*
SCC CV 0.030
(<= 0.040)
IN
MUN AVG 1212
-3%
(1250 +/- 10%)
IN
MUN CV 0.008
 

In the check sample results printout the components for which tolerance values are specified will show the tolerance range and "IN" or "OUT" flags. Components with a standard value but no tolerance specified will show only the deviation and the standard value.

 
clear
Locations        
Columbus
  Phone: 614-545-0460
  Toll Free: 800-344-6446
  Fax: 614-545-0468
Columbus Street Address:
  DHI Cooperative, Inc.
  1224 Alton-Darby Creek Rd.
  Suite A
  Columbus, OH 43228
Columbus Mailing Address:
  DHI Cooperative, Inc.
  P.O. Box 28168
  Columbus, OH 43228
Wooster
  Phone: 330-264-0411
  Toll Free: 800-856-2817
  Fax: 330-264-0733
Wooster Address:
  DHI Cooperative, Inc.
  Wooster Lab
  7849 E. Lincolnway
  Apple Creek, OH 44606
Home  I  Services  I  Lab System  I  Links  I  BioPRYN® Pregnancy Blood Test Kits  I  Agricltural Technology  I  PcDart  I  PocketDairy  I  DairyMetrics  I  WebReports  I  Contact Us


Website Design by EDJE technologies